Omron Healthcare Blood Pressure Monitor HEM 907 User Manual

INSTRUCTION MANUAL  
SYS  
mm  
Hg  
DIA  
OMRON  
mmHg  
PU  
L
SE  
Digital Automatic Blood Pressure Monitor  
O
N
/
OF  
BPM  
HID  
F
E
STA  
RT  
N
P
1
40  
-
SE  
T
Model  
1
80  
1
00  
MOD  
E
SI  
D
A
VG.  
N
G
A
E
220  
LE  
V
F
L
s
t
G
A
/
T
n
.
I
d
/
O
/
1
M
2
AN  
CHE  
3
r
d
AU  
U
T
O
.
260  
280  
CK  
ST  
O
P
X
A
M
INDEX  
HEM-907  
Table of Contents  
Thank you very much for purchasing the  
OMRON Digital Automatic Blood Pressure  
Monitor.  
Intended use................................................. 2  
Exemptions................................................... 3  
Names and Functions of the Parts...............4  
Notes on Safety............................................ 8  
Components of the Product ....................... 12  
Options....................................................... 12  
How to Apply the Arm Cuff......................... 13  
How to Use the Power Source  
(AC Adapter)........................................... 14  
How to Measure Blood Pressure ............... 15  
SINGLE Mode........................................ 16  
AVG. Mode............................................. 17  
MANU. Mode.......................................... 18  
CHECK Mode......................................... 19  
Installation and Replacement of  
Requests from OMRON to the  
operators and the persons  
responsible for maintenance  
Please read thoroughly the “Notes on Safety”  
of this Instruction Manual before using this unit  
so that you can use it safely and correctly after  
suf[cient understanding.  
After reading this Instruction Manual, please  
keep it near the unit all the times for future  
reference.  
Battery Pack...........................................20  
How to Clean the Unit after Use................. 21  
List of Error Codes .....................................22  
Troubleshooting..........................................23  
Speci[cations............................................. 24  
= Read the instruction manual carefully  
IM-HEM-907-E7-03-01/2012  
5329412-4C  
1
 
Exemptions  
OMRON will not bear any responsibilities on the following matters.  
1. When a problem or damage occurs caused by the maintenance and/or repair con-  
ducted by a person other than OMRON or the dealer speci[ed by OMRON  
2. The problem or damage of OMRON product caused by the product of other manu-  
facturer not delivered by OMRON  
3. The problem and damage caused by the maintenance and/or repair using the repair  
parts not speci[ed by OMRON  
4. The problem and damage caused by the results not observing the Notes on Safety  
or the operational method mentioned in this Instruction Manual  
5. Under the circumstances not within the operating conditions of this unit including the  
power source or the setting environment mentioned in this Instruction Manual  
6. The problem and damage caused by the result(s) of remodeling or improper repair  
of this product  
7. The problem and damage caused by act of god such as [re, earthquake, \ood, or  
lightening  
1. The contents of this Instruction Manual may be changed without prior notice.  
2. We have thoroughly reviewed the contents of this Instruction Manual.  
However, if an inadequate description or error is found, please let us know.  
3. It is prohibited to copy a part of or the entire Instruction Manual without getting  
OMRON’s permission. Unless this Instruction Manual is used by an individual (com-  
pany), it cannot be used without getting OMRON’s permission from the standpoint  
of the Copyright Law.  
3
 
Names and Functions of the Parts  
Main unit  
1. LCD display  
SYS  
6. ON/OFF  
(power) Button  
2. HIDE (non-display) Button  
mmHg  
DIA  
7. START Button  
mmHg  
PULSE  
ON/OFF  
BPM  
DC8V  
AC  
8. DEFLATION  
(de\ation  
control) Button  
HIDE  
START  
P
-
SET  
14  
3. DC jack  
0
1
8
0
10  
0
MODE  
SINGLE  
VG.  
D
A
A
E
220  
F
V
L
G
A
T
9. Arm Cuff  
Connector  
.
I
d
/
O
N
1
s
t
/
MANU.  
CHECK  
2
n
/
3
r
d
AUTO  
260  
280  
4. P-SET (pressure  
setting) Volume  
ST  
OP  
10. STOP Button  
5. MODE Selector  
4
 
Names and Functions of the Parts  
Ready to Measure  
Display  
Pulse level / Number of  
irregular pulse waves  
Number of irregular pulse waves:  
The pulse wave having a difference  
of more than 25% from the average  
interval of pulse waves is determined  
as irregular. Number of irregular pulse  
waves are displayed up to three times.  
SYS  
Systolic blood  
pressure  
mmHg  
Example of display  
No. of arrhythmias  
DIA  
1
2
3
Diastolic blood  
pressure  
mmHg  
BPM  
PULSE  
AC  
Battery level  
Displays  
for the usable level.  
for the low level.  
HIDE  
Displays  
Displays  
for the unusable level.  
Contents when the AVG.  
Mode is selected  
Pulse rate  
AVG. : Mean value  
1st : First measurement  
2nd : Second measurement  
3rd : Third measurement  
Charging  
Displays  
when the battery  
pack is being charged.  
External power source  
Displays AC when the unit is con-  
nected to the AC adapter.  
5
 
Names and Functions of the Parts  
Explanation of Functions  
-
P SET  
(1) Pressure setting function  
In\ation can be set by AUTO (automatic setting) or the target value.  
AUTO (automatic setting): In the SINGLE (single measurement), AVG. (average), and MANU. (auscultation) Modes, the  
monitor estimates the systolic blood pressure value during in\ation and automatically in\ates to the proper value.  
Target value setting: The monitor in\ates to the target value.Pressure value is set to 30 to 40 mmHg above the expected  
systolic pressure.  
HIDE  
(2) Non-display function  
ON/OFF  
A function not to display the results of measured blood pressures.  
However, the pressure values during the measurement are displayed.  
START  
HIDE  
This function can be used when SINGLE (single measurement) or AVG.  
(average) Mode is selected.  
-
P
SET  
MODE  
SINGLE  
140  
180  
AVG.  
MANU.  
CHECK  
220  
260  
100  
AUTO  
280  
(3) Manual de\ation control function  
ON/OFF  
A function to accelerate the de\ation speed by pushing a DEFLATION Button  
during de\ation.  
START  
HIDE  
This function can be used when MANU. (auscultation) Mode is selected.  
A
/
-
P
SET  
180  
MODE  
SINGLE  
140  
AVG.  
MANU.  
CHECK  
220  
260  
100  
STOP  
AUTO  
280  
6
 
Names and Functions of the Parts  
(4) Setting mode change function  
You can set the number of times to measure, the waiting time until the start of measurement, and the interval between  
measurements.  
Items to set  
Set value  
F1  
F2  
F3  
Number of measurements  
Waiting time until the start of measurement  
Measurement interval time  
2 times, 3 times  
0 sec, 3 min, 5 min, 10 min.  
5 sec, 30 sec, 1 min, 2 min, 3 min.  
Procedure to change the set values  
ON/OFF  
1) When the power is off, press the ON/OFF Button for more than three  
seconds while holding the START Button and change the mode to the  
Setting Change Mode.  
START  
HIDE  
A
/
-
P
SET  
180  
MODE  
SINGLE  
140  
AVG.  
MANU.  
CHECK  
220  
260  
2) Press the START Button to select the item from F1 to F3.  
3) Press the DEFLATION Button to change the set values.  
100  
STOP  
AUTO  
280  
List of Functions for each Mode  
Measurement  
Single measure-  
ment (SINGLE)  
Average  
(AVG.)  
Auscultation  
(MANU.)  
Check  
(CHECK)  
Mode  
Function  
Pressure setting  
function  
Non-display  
function  
Manual de\ation  
control function  
Setting mode  
change function  
7
 
Notes on Safety  
The warning signs and the sample icons shown here are listed for you to use the product safely and correctly as well as  
tTohpereicvoennst athnedrmisekaannindgtsheardeaamsafgoellotow.you and others from happening.  
Warning sign  
Contents  
Indicates matters in which death or sever bodily damage may arise as a result of incorrect  
handling.  
Warning  
Indicates matters in which bodily harm or material damage* may arise as a result of incorrect  
handling.  
Caution  
* Material damage refers to a wide range of damage involving your house, household goods, domestic animals, and pets.  
Examples of signs  
The icon indicates caution (including warning and danger). Matters involving actual caution are  
indicated by text or pictures in or near . The left icon refers to “caution for ignition”.  
The  
icon indicates prohibitions (what you cannot do). Matters involving actual prohibitions are indi-  
. The left icon refers to “prohibition to disassemble”.  
cated by text or pictures in or near  
The  
icon indicates something that is compulsory (always follow). Matters involving actual compul-  
sory actions are indicated by text or pictures in or near  
source plug”.  
. The left icon refers to “pulling the power  
Warning  
If any abnormal matter occurs during the measurement such as in\ation does not stop, remove the arm  
cuff or pull out the air tube from the main unit.  
You may suffer peripheral neuropathy.  
Do not wrap the arm cuff over the arm to which intravenous injection or transfusion is being conducted.  
Intravenous injection or transfusion is impossible.  
Do not use the unit in the place where in\ammable gas, such as highly in\ammable anesthetic, may be  
generated or in a high pressure oxygen room or an oxygen tent.  
It may cause ignition and explosion.  
Do not touch the AC adapter with wet hands.  
You may suffer electric shock.  
8
 
Notes on Safety  
Caution  
When you are not going to use the unit for a long period of time, be sure to remove the AC adapter  
from the electric outlet.  
You may suffer electric shock, or electric leak or [re may arise because of deteriorated insulation.  
Be sure to remove the AC adapter from the electric outlet when installing, removing, or cleaning the  
part.  
You may suffer electric shock or injure yourself.  
Pull the AC adapter from the electric outlet when cleaning the unit.  
You may suffer electric shock.  
In the following case, con[rm the measurement with the stethoscope.  
(1) When an irregular pulse wave is displayed  
· An error may be displayed when an external vibration is applied or the patient moves his/her body.  
(2) When an error is generated or the measured value is doubtful  
· An error may be displayed when an external vibration is applied or the patient moves his/her body.  
· An error may be generated in the measured value or the blood pressure may not be able to measure for the  
patient having peripheral circulatory disturbance or extreme hypotension.  
· An error may be generated in the measured value or the blood pressure may not be able to measure for the  
patient having short variation of blood pressure such as arrhythmia.  
After cleaning the unit, dry it well, then insert the AC adapter to the electric outlet.  
You may suffer electric shock.  
Insert the AC adapter to the electric outlet as far as it goes.  
Wipe off the dust on the AC adapter.  
You may suffer electric shock, or there may be short circuit or [re ignition.  
If you [nd any trouble with this unit, immediately stop using it, turn off the power, pull out the AC  
adapter from the electric outlet, then contact the repair department.  
You may suffer electric shock, or there may be short circuit or [re ignition.  
Do not disassemble or remodel the unit.  
You may suffer electric shock, or there may be short circuit or [re ignition.  
Do not use the unit to the patient using a pump oxygenator.  
The unit can not measure correctly. The unit cannot measure when it cannot detect any beat.  
Do not use the AC adapter and the battery pack not speci[ed for this unit.  
It may cause [re or electric shock.  
Do not use a portable phone near the unit.  
The unit may malfunction.  
Do not install the parts and/or instrument not speci[ed for this unit.  
It may cause damage to the unit.  
Do not use the damaged power cord or AC adapter, or loose electric outlet.  
You may suffer electric shock, or there may be short circuit or [re ignition.  
Do not damage the power cord of the AC adapter by placing heavy thing on it or pinching it, or modify,  
forcibly bend, pull, twist, or bundle it.  
You may suffer electric shock or there may be electric leak and [re because of deteriorated insulation.  
9
 
Notes on Safety  
Caution  
Be sure to use the power supply of 230 VAC.  
It may cause [re or electric shock.  
Do not share an electric outlet with other unit or electric appliance.  
It may cause [re or electric leak.  
Do not pull the power cord when pulling out the AC adapter from the electric outlet.  
The power cord will be disconnected or shorted and may cause [re or electric shock.  
Do not install or store the unit where it may be sprayed with water or medication.  
You may suffer electric shock.  
Requests from OMRON  
Do not place or put anything on the unit.  
Do not apply strong shock to or drop the unit.  
When you move the unit, be sure to turn off the power and remove the connected AC adapter.  
When using this unit concurrently with other unit, be sure to read the Instruction Manual of the other unit to be used  
concurrently and understand the warnings and cautions before use.  
Be sure to con[rm that all cords and tubes are connected correctly and completely before use.  
Inspect the Buttons and so on to con[rm that the unit operates normally before use.  
When using a battery pack, be sure to con[rm that the voltage is suf[cient before use. If you are not going to use the  
unit for a long period of time, remove the battery pack.  
Do not in\ate the arm cuff without being wrapped over the arm.  
Do not use a damaged arm cuff.  
Do not use the unit in a vehicle.  
Be sure to monitor all the time that the unit and the patient are not abnormal.  
Please keep the unit out of the reach of patient.  
Clean and store the unit and accessories after each use.  
Do not clean the unit with gasoline, thinner, or high concentration alcohol.  
Do not disinfect the unit by autoclave or gas sterilization (EOG, formaldehyde, or high concentration ozone.)  
Do not store the unit in the following places.  
· Under the direct sunshine  
· Dusty or salty environment  
· Places which slope, vibrate, and/or are prone to shocks  
· Storage of chemicals or where gas may be generated  
· Under high temperature and high humidity  
10  
 
Notes on Safety  
Maintenance  
1. Be sure to inspect the unit on regular basis. Check accuracy on a 1 year interval.  
2. If the unit has not been used for a while, be sure to con[rm that the unit operates normally and safely before use  
Durability  
The durability of this unit is [ve years. (Arm cuff is a consumable.)  
Measures to take at the time of trouble or accident  
If a device error (Er9) occurs, take the following procedure promptly.  
1. Remove the arm cuff from the patient’s arm.  
2. Turn off the power and pull the AC adapter from the electric outlet. If an optional battery pack is used, remove it also.  
3. Display “Trouble” on the unit so that it cannot be used.  
4. Contact the dealer where you purchased the unit or the nearest OMRON dealer  
11  
 
Components of the Product  
Main unit  
Accessories  
Medium size arm cuff (with built-in air bag)  
Model: HEM-9CMC  
Instruction Manual  
(with guarantee card)  
Art.no.: 4928640-0  
Air tube (1 m)  
Art.no.: 4928647-8  
Applicable arm circumference: 22 to 32 cm  
MIN  
RANGE  
MAX  
ON/OFF  
HIDE  
START  
P
-
S
E
T
14  
0
1
80  
10  
0
MODE  
S
D
E
IN  
AV  
2
G
L
E
A
V
G
2
0
G
F
.
L
A
T
n
.
I
O
d
/
1
N
/
3
s
t
MA  
C
/
2
rd  
A
U
N
U
TO  
.
260  
280  
HE  
AC adapter  
Battery pack  
(48H907N-E)  
Model: HEM-9BAT  
Art.no.: 1098391-0  
C
K
ST  
OP  
(80230H907S)  
Model: HEM-9ADAP  
Art.no.: 1098200-0  
Options  
Large size arm cuff  
Model: HEM-9CLC  
Art.no.: 4928654-0  
Applicable arm  
Medium size arm cuff  
Small size adult arm cuff  
Model: HEM-9CSC  
Art.no.: 4928639-7  
Applicable arm  
Model: HEM-9CMC  
Art.no.: 4928640-0  
Applicable arm  
circumference:  
circumference:  
circumference:  
32 to 42 cm  
22 to 32 cm  
17 to 22 cm  
MIN  
RANGE  
MIN  
MAX  
RANGE  
MAX  
MIN  
RANGE  
MAX  
Air tube (1.3 m)  
Model: HEM-9T1.3  
Art.no.: 4928648-6  
Stand exclusive  
for this unit  
Model: HEM-9ST  
Art.no.: 4928649-4  
Wall-hanging kit  
Model: HEM-9WM  
Art.no.: 4928650-8  
Pole-mounting kit  
Model: HEM-9PM  
Art.no.: 4928651-6  
12  
 
How to Apply the Arm Cuff  
Requests from OMRON  
Warning  
Do not use the unit to the patient using a pump oxy-  
Do not apply the arm cuff during the intravenous  
injection or transfusion.  
gDeonnaototri.n\ate the unit without applying the arm cuff.  
Intravenous injection or transfusion is impossible.  
Do not use the damaged cuff.  
MIN  
RANGE  
MAX  
1. Select the arm cuff according to the arm circumference of the patient.  
Arm circumference  
17 - 22 cm  
22 - 32 cm  
32 - 42 cm  
Name of the arm cuff  
Small size adult arm cuff (option)  
Adult arm cuff  
MIN  
RANGE  
MAX  
Large size adult arm cuff (option)  
MIN  
RANGE  
MAX  
Be sure to use the arm cuff suitable for the size of the patient’s arm.  
If an arm cuff not suitable for the size of the arm is used, blood pressure  
may not be measured correctly.  
Cuff side  
Air tube  
side  
2. Securely connect the air tube.  
If you connect the attached 1m air tube, the air tube can be used with the  
entire length of 1.2 m.  
3. Prepare the patient to be ready to wrap the arm cuff.  
Wrap the arm cuff over the right upper arm’s skin as a rule.  
If the arm cuff is wrapped over a thick cloth or by rolling up the jacket or shirt, blood pressure may not be measured  
correctly.  
4. Place the right hand of  
the patient with the palm  
of hand facing upward.  
5. Place the arm cuff over the patient’s  
arm by matching the mark indicating  
the artery position  
ART.  
to the brachial artery.  
6. Wrap the arm cuff snugly using both hands and securely fasten it with the Velcro tape. At this time, the lower  
edge of the arm cuff must be placed 1 to 2 cm above the inner side of elbow joint.  
If the arm cuff is not wrapped snugly, patient may feel numbness on his/her arm, or blood pressure may not be  
mTheeasaurmredcucfofrwreractplyp.ed diagonally along the shape of the arm does not affect the measurement value.  
1-2 cm  
13  
 
How to Apply the Arm Cuff  
(cont.)  
7. Adjust the level of the arm cuff to the level of the heart.  
Keep the level of the arm cuff at the same level as the heart during the measure-  
ment.  
How to use the Power Source  
(AC adapter exclusive for this unit)  
Connect the AC adapter to the DC jack of the main unit  
[1] and the electric outlet [2].  
Warning  
Do not use the unit in the place where in\amma-  
ble gas, such as highly in\ammable anesthetic,  
may be generated, or in the high pressure  
oxygen room or the oxygen tent.  
It may cause ignition and explosion.  
Do not touch the AC adapter with wet hands.  
You may suffer electric shock.  
[2]  
[1]  
Caution  
D
C
8
V
Be sure to use the power supply of 230 VAC.  
It may cause [re or electric shock.  
Do not install or store the unit where it may be  
sprayed with water or medication.  
You may suffer electric shock.  
When using an optional battery pack, the AC adapter  
functions as the charger also.  
Requests from OMRON  
If this unit is used concurrently with other unit, be sure  
to read the Instruction Manual of the other unit to be  
used concurrently and understand the warnings and  
cInasupteiocntsthbeefBourettounsse.to con[rm that the unit operates  
normally before use.  
14  
 
How to Measure Blood Pressure  
Warning  
Caution  
If any abnormal matter occurs during the mea-  
surement such as the in\ation does not stop,  
remove the arm cuff or pull the air tube from the  
main unit.  
In the following case, con[rm the measurement  
with the stethoscope.  
(1) When an irregular pulse wave is displayed  
· An error may be displayed when an external vibra-  
tion is applied or the patient moves his/her body.  
(2) When an error is generated or the measured value  
is doubtful  
· An error may be displayed when an external vibra-  
tion is applied or the patient moves his/her body.  
· An error may be generated in the measured valueor  
the blood pressure may not be able to measure for  
the patient having peripheral circulatory disturbance  
or extreme hypotension.  
You may suffer peripheral neuropathy.  
Requests from OMRON  
Be sure that the patient should not touch the unit.  
Be sure to monitor all the time that the unit and the  
patient are not abnormal.  
Do not use the unit in a vehicle.  
· An error may be generated in the measured valueor  
the blood pressure may not be able to measure for  
the patient having short variation of blood pressure  
such as arrhythmia.  
Do not use a portable phone near the unit.  
The unit may malfunction.  
List of Measurement Modes  
Refer to Page  
To measure  
only once  
SINGLE Mode  
16.  
To measure three  
times  
Refer to Page  
AVG. Mode  
(or two times)  
consecutively  
17.  
To measure  
by using a  
Refer to Page  
MANU. Mode  
18.  
stethoscope  
To con[rm the  
Refer to Page  
accuracy of  
CHECK Mode  
19.  
pressure display  
15  
 
How to Measure Blood Pressure  
(in SINGLE Mode)  
1. Wrap the arm cuff over the  
patient’s arm.  
2. Push the ON/OFF (power)  
Button to turn on the power.  
ON/OFF  
START  
A
HIDE  
/
n
-
P
SET  
180  
MODE  
SINGLE  
140  
AVG.  
MANU.  
CHECK  
220  
260  
100  
STOP  
AUTO  
280  
3. Set the MODE Selector to  
“SINGLE”.  
4. Set the P-SET (pressure set-  
ting) Volume to “AUTO” or the  
target in\ation value.  
ON/OFF  
ON/OFF  
START  
A
START  
A
HIDE  
HIDE  
When setting the P-SET to  
/
n
/
n
-
P
SET  
180  
MODE  
SINGLE  
-
P
SET  
180  
MODE  
SINGLE  
140  
140  
AVG.  
MANU.  
CHECK  
AVG.  
MANU.  
CHECK  
220  
260  
220  
60  
100  
“AUTO”, turn the Volume  
STOP  
STOP  
AUTO  
280  
AUTO  
counterclockwise as far as it  
goes until you can hear the  
cWlihckensothuendsy. stolic pressure is  
expected to exceed 220 mmHg,  
proper in\ation may not be pos-  
sible in “AUTO” setting.  
Set the target in\ation value to  
30 to 40 mmHg higher than the  
expected systolic pressure.  
6. Measured results are dis-  
played.  
5. Push the START Button to start  
the measurement.  
SYS  
ON/OFF  
mmHg  
DIA  
START  
A
If the in\ation is determined  
HIDE  
insuf[cient, the unit may start  
/
n
-
mmHg  
BPM  
P
SET  
180  
MODE  
SINGLE  
140  
PULSE  
AVG.  
MANU.  
CHECK  
220  
100  
iInf \yaotuiownaangtatoinsatouptommeaaticsuarlley-.  
ment, push the STOP Button.  
The unit de\ates fast.  
STOP  
260  
280  
AUTO  
HIDE  
7. Push the ON/OFF (power)  
Button to turn off the power.  
ON/OFF  
START  
A
HIDE  
/
n
-
P
SET  
180  
MODE  
SINGLE  
140  
AVG.  
MANU.  
CHECK  
220  
100  
STOP  
260  
280  
AUTO  
16  
 
How to Measure Blood Pressure  
(in AVG. Mode)  
1. Wrap the arm cuff over the  
patient’s arm.  
2. Push the ON/OFF (power)  
Button to turn on the power.  
ON/OFF  
START  
HIDE  
A
/
n
-
P
SET  
180  
MODE  
SINGLE  
140  
AVG.  
MANU.  
CHECK  
220  
260  
100  
STOP  
AUTO  
280  
4. Set the P-SET (pressure set-  
ting) Volume to “AUTO” or the  
target in\ation value.  
3. Set the MODE Selector to  
“AVG.”.  
ON/OFF  
ON/OFF  
START  
A
START  
A
HIDE  
HIDE  
When setting the P-SET to  
“AUTO”, turn the Volume coun-  
terclockwise as far as it goes  
until you can hear the click  
/
n
/
n
-
P
SET  
180  
MODE  
SINGLE  
-
P
SET  
180  
MODE  
SINGLE  
140  
140  
AVG.  
MANU.  
CHECK  
AVG.  
ANU.  
ECK  
220  
260  
220  
260  
100  
STOP  
STOP  
AUTO  
0  
AUTO  
280  
sWohuennd.the systolic pressure is  
expected to exceed 220 mmHg,  
proper in\ation may not be pos-  
sible in “AUTO” setting.  
Set the target in\ation value to  
30 to 40 mmHg higher than the  
expected systolic pressure.  
6. Measured results are  
displayed.  
5. Push the START Button to start  
the measurement.  
SYS  
ON/OFF  
mmHg  
After the measurement is  
completed, average values are  
displayed.  
Set the unit to the start time of  
the [rst measurement, then the  
unit starts measurement auto-  
matically.  
DIA  
START  
A
HIDE  
/
n
mmHg  
BPM  
-
P
SET  
180  
MODE  
SINGLE  
PULSE  
140  
AVG.  
MANU.  
CHECK  
220  
100  
Each time a DEFLATION  
(de\ation control) Button is  
pushed, the measurement  
results for each time are  
displayed.  
STOP  
260  
280  
AUTO  
After displaying the results of  
[rst measurement, the unit will  
measure blood pressure in the  
set number of times automati-  
HIDE  
cFaolrlythbeystaekttiinnggaonf tihnetenrvuaml.ber of  
measurements, the waiting time  
before start of measurement,  
and the interval time, refer to  
Page 7.  
7. Push the ON/OFF (power)  
Button to turn off the power.  
To stop the measurement during the measurement:  
Push the STOP Button.  
To start the stopped measurement again:  
Push the START Button.  
ON/OFF  
START  
A
HIDE  
/
n
-
P
SET  
180  
MODE  
SINGLE  
140  
AVG.  
MANU.  
CHECK  
220  
100  
STOP  
260  
280  
AUTO  
17  
 
How to Measure Blood Pressure  
(in MANU. Mode)  
1. Wrap the arm cuff over the  
patient’s arm.  
2. Place the stethoscope on the  
patient’s arm.  
ON/OFF  
START  
A
/
d
S
40  
E  
NLE  
1
U.  
HECK  
0
2
2
STOP  
AUTO  
280  
4. Set the MODE Selector to  
“MANU.”.  
3. Push the ON/OFF (power)  
Button to turn on the power.  
ON/OFF  
ON/OFF  
START  
A
START  
A
HIDE  
HIDE  
/
n
/
n
-
P
SET  
180  
MODE  
SINGLE  
-
P
SET  
180  
MODE  
SINGLE  
140  
140  
AV
MANU.  
CHECK  
AVG.  
MANU.  
CHECK  
220  
100  
220  
260  
100  
STOP  
STOP  
AUTO  
AUTO  
280  
6. Push the START Button to  
start the measurement.  
5. Set the P-SET (pressure set-  
ting) Volume to “AUTO” or the  
target in\ation value.  
ON/OFF  
ON/OFF  
In the “AUTO” setting, the unit  
START  
A
START  
A
HIDE  
HIDE  
stops in\ation at the pres-  
When setting the P-SET to  
“AUTO”, turn the Volume coun-  
terclockwise as far as it goes un-  
tWil hyoeun cthaenshyesatorlticheprceliscskusreouisnd.  
expected to exceed 220 mmHg,  
proper in\ation may not be pos-  
sible in “AUTO” setting.  
/
n
/
n
-
P
SET  
180  
MODE  
SINGLE  
-
P
SET  
180  
MODE  
SINGLE  
sure considered to be 30 to  
40 mmHg above the expected  
systolic pressure, then start  
140  
140  
AVG.  
MANU.  
CHECK  
AVG.  
MANU.  
CHECK  
220  
100  
220  
60  
STOP  
STOP  
260  
280  
AUTO  
AUTO  
dTeh\eautinoint .can be re-in\ated  
only while the START Button is  
pTuhsehperde.ssure display is syn-  
Set the target in\ation value to  
30 to 40 mmHg higher than the  
expected systolic pressure.  
chronized with the pulse rate  
aEnadchretnimeewethde. DEFLATION  
(de\ation control) Button is  
pushed, the unit de\ates by  
5W~h1e0nmymouHpgr.ess the STOP But-  
ton, the unit de\ates fast and  
ends the operation.  
7. Push the ON/OFF (power)  
Button to turn off the power.  
ON/OFF  
START  
A
HIDE  
/
n
-
P
SET  
180  
MODE  
SINGLE  
140  
AVG.  
MANU.  
CHECK  
220  
260  
100  
STOP  
AUTO  
280  
18  
 
How to Measure Blood Pressure  
Display of pressure can be con[rmed by the CHECK Mode (yearly procedure).  
What you need to prepare  
(in CHECK Mode)  
(1) Well-adjusted reference pressure monitor (including rubber ball),  
(2) T-shaped tube,  
(3) Two rubber tubes, and (4) something in the cylindrical shape to wrap the arm cuff such as the tea container (sturdy one  
that will not break or deform with pressure)  
* Measurement error may occur even with the mercury blood pressure monitor because of insuf[cient amount of mercury  
in the mercury, dirty glass tube, or clogged glass tube cap.  
How to use the blood pressure monitor  
1. Connect the mercury blood  
pressure monitor, rubber  
ball, and this unit with the  
S
Y
S
T-shaped tube as shown in  
the Figure.  
mmH  
g
DIA  
ULS  
2. Push the ON/OFF (power)  
Button to turn on the  
power.  
mmHg  
P
E
O
N
/O  
F
F
B
PM  
AC  
3. Set the MODE Selector to  
H
IDE  
S
T
AR  
T
“CHECK”.  
P
0
-
SET  
14  
1
8
0
MOD  
SING  
D
G
10  
E
E
.
/
1
s
0
A
V
F
L
t
A
/
2
T
n
I
d
O
N
AV  
LE  
G
.
220  
/
3r  
d
4. Tightly wrap the arm cuff  
over a sturdy cylindrical  
object.  
M
A
NU  
CHEC  
.
AU  
260  
280  
K
T
O
ST  
OP  
5. Close the air release valve  
of rubber ball and in\ate  
the monitor to a certain  
value (pressure to be  
checked).  
6. Compare the value displayed on this monitor and that on the mercury.  
7. Open the air release valve of rubber ball to de\ate.  
CHECK result  
(1) In the measuring rule, the test tolerance of blood pressure monitor (accuracy of pressure when shipped from the  
factory) should not exceed 3 mmHg.  
(2) If the CHECK result is +/- 3 mmHg or above, contact the dealer where you purchased this unit or the nearest OMRON  
dealer.  
19  
 
Installation and Replacement of Battery Pack  
Warning  
Caution  
Do not disassemble or modify the battery pack.  
Do not throw the battery pack into [re or heat it.  
Do not charge the battery pack by using equip-  
ment other than this unit.  
Do not short the polarities of battery using metal  
object such as the wire.  
If the \uid in the battery is stained on your skin  
or cloth, immediately wash off the \uid with  
water.  
If the \uid in the battery pack gets into your  
eye, wash the eye with suf[cient water without  
rubbing the eye. Then immediately consult the  
doctor for treatment.  
You may suffer injury, or battery \uid may leak,  
or battery may heat, ignite [re, or explode.  
You may suffer injury, or battery \uid may leak,  
or battery may heat, ignite [re, or explode.  
1. Remove the battery cover on the back of the  
main unit.  
2. Installation: To install the battery pack, connect  
the battery pack connector to the connector in the  
battery cover.  
Replacement: Remove the battery pack from the  
connector and replace with a new one.  
3. Install the battery cover and fasten it with  
screws.  
Battery life  
You can use the unit for approximately three hundred measurements.  
If a  
mark appears frequently even after the battery is charged, replace the battery.  
Approximate battery life is two years.  
Charging time  
After inserting the AC adapter, the battery will start to charge automatically in approximately [ve seconds.  
While the battery is being charged, a  
mark turns on.  
The battery can be charged in approximately twelve hours.  
Battery low  
Even after a  
If a  
mark starts to \ash, you can use the battery for twenty to thirty measurements. However it is  
better to charge the battery as soon as possible.  
mark is displayed, the battery is low. Please charge the battery.  
20  
 
How to Clean the Unit after Use  
Requests from OMRON  
Caution  
Pull the AC adapter from the electric outlet when  
cleaning the unit.  
You may suffer electric shock.  
Do not clean the unit with gasoline, thinner, or high  
concentration alcohol.  
After cleaning the unit, dry it well, then insert the  
Do not sterilize the unit by autoclave or gas sterilization  
AC adapter to the electric outlet.  
(EOG, formaldehyde, or high concentration ozone.)  
You may suffer electric shock.  
1. Clean the blood pressure monitor,the cuff and the  
cables using a cloth dampened with water,diluted  
disinfectant alcohol or diluted mild detergent.  
To disinfect, wipe the unit with a soft cloth squeezed  
well after moistened with the following disinfectant.  
Benzalkonium chloride  
Chlorhexizin  
Amphoteric surface active agent  
0.01 - 0.2 w/v%  
0.05 - 0.5 w/v%  
0.01 - 0.2 w/v%  
When cleaning or disinfecting the unit, do not wipe the  
ON/OFF  
AC adapter.  
H
ID  
E
S
T
A
R
T
P
0
-
S
ET  
14  
2. Then wipe the unit with a soft dry cloth.  
1
8
0
MODE  
1
00  
D
A
E
F
S
IN  
V
L
A
V
G
L
G
A
T
t
2
G
E
.
/
1
I
O
/
N
3
20  
.
s
/
2
n
d
r
d
M
A
N
C
H
U
.
A
U
T
26  
E
O
C
0
K
2
8
0
ST  
OP  
21  
 
List of Error Codes  
Error code  
Cause  
How to correct  
In\ation error  
If the connector of rubber tube connecting the  
arm cuff and the main unit is loose, connect them  
correctly, then measure.  
Er1  
When the pressure does not exceed 15  
mmHg after operating a pump for 15 seconds  
fWrohmenthtehestianr\taotifoinn\daoteiosnnot reach the set  
pressure of the arm cuff within the speci[ed  
time after starting the in\ation  
If the rubber tube connecting the arm cuff and the  
main unit is bent, stretch it, then measure.  
If the arm cuff is wrapped loosely, wrap it correctly,  
then measure.  
De\ation error  
Er2  
When the de\ation speed is too fast during  
tWhehemnetahseudreem\aetniotn speed is too slow during  
tWhehemnetahseumreemaesnutrement does not end  
within the speci[ed time after starting the  
mWehaesnutrheempernetssure is applied more than 5  
minutes  
Overpressure error  
If the rubber tube connecting the arm cuff and the  
main unit is bent, stretch it, then measure.  
Er3  
Er4  
The arm cuff pressure exceeded 299 mmHg.  
Insuf[cient in\ation error  
If the measurement is made by setting P-SET to  
“AUTO”, ask the patient not move his/her arm or  
body during the in\ation.  
Blood pressure could not be measured due  
to insuf[cient in\ation after the measurement  
ends.  
When setting the P-SET to “AUTO”, turn the Volume  
counterclockwise as far as it goes until you can hear  
the click sound.  
Set the P-SET to 30 to 40 mmHg higher than the  
systolic pressure of the patient, then measure.  
Indeterminable blood pressure error  
Blood pressure could not be measured even  
when the arm cuff pressure reached the  
speci[ed pressure.  
If the arm cuff is wrapped loosely, wrap it correctly,  
Er5  
Er6  
then measure.  
Pulse wave small error  
Pulse wave was too small.  
If the arm cuff is wrapped loosely, wrap it correctly,  
tDhoennomtewarsaupreth.e arm cuff over a thick cloth or by  
rolling up the jacket or shirt.  
Measurement error  
Check if the patient has an arrhythmia.  
Er7  
Er8  
Er9  
Relationship between systolic and diastolic  
Ask the patient not to move his/her arm or body  
pressures was abnormal.  
during the measurement.  
Pulse rate error  
Pule rate did not stay within the range of 30  
to 199 beats/min.  
Device error  
Contact the dealer where you purchased the unit or  
the nearest OMRON dealer.  
There is a problem with the main unit of the  
blood pressure monitor.  
22  
 
Troubleshooting  
If a trouble occurs while using the unit, please check the following.  
Problem  
What to inspect  
How to correct  
Wrap the arm cuff correctly, and measure  
again.  
Is the arm cuff wrapped correctly?  
Ask the patient not to move his/her arm or  
body during measurement, and measure  
again.  
Is the patient moving his/her arm or body  
The unit in\ates to abnormally during in\ation?  
high (low) pressure.  
Does this trouble occur with a speci[c  
patient?  
Check the measurement by the stetho-  
scope.  
For patients having arrhythmia, the unit  
may not in\ate properly  
After checking the patient with the stetho-  
scope as needed, take a proper mea-  
sure in accordance with the “list of error  
codes”.  
First of all, check the patient’s condition.  
Ask the patient not to move his/her arm or  
body during measurement, and measure  
again.  
Is the patient moving his/her arm or body  
during measurement?  
The unit cannot measure  
blood pressure.  
Does the patient have an arrhythmia?  
Check the patient with a stethoscope.  
The blood pressure values are  
extremely high (low).  
Select the arm cuff in accordance to the  
circumference of the patient’s arm, wrap it  
correctly, then measure again.  
Is the size of the arm cuff and the wrapping  
of arm cuff correct?  
Is the level of the brachium to which the  
arm cuff is wrapped at the same level as  
the heart?  
Keep the level of the brachium to which  
the arm cuff is wrapped at the same level  
as the heart, then measure again.  
23  
 
Speci[cations  
Name  
:
:
:
:
:
OMRON Digital Automatic Blood Pressure Monitor HEM-907  
HEM-907-E / HEM-907-E7  
Model  
Display  
Digital display  
Measurement  
Measurement Range  
Oscillometric method  
Pressure; 0 to 299 mmHg  
Pulse rate: 30 to 199 beats/min  
Pressure; Within +/-3 mmHg  
Accuracy  
:
Pulse rate; Within +/-5% of reading  
Automatic in\ation with pumping  
Automatic de\ation by electromagnetic control valve  
Automatic rapid air release by electromagnetic control valve  
AC adapter (230 VAC, 50 Hz. 20 VA)  
or battery pack (4.8 VDC, 6 W)  
Class II B type  
In\ation  
:
:
:
:
De\ation  
Air Release  
Power supply  
Electric Shock Protection Method  
:
Operating Temperature and Humidity : 10 to 40°C, 30 to 85% RH  
Weight of Main Unit  
:
:
:
Approximately 910 g  
External Dimensions  
139 (W) x 203 (H) x 131 (D) mm  
Also included in the package  
Adult arm cuff (with built-in air bag), AC adapter, air tube (1 m), Instruction  
Manual (with guarantee card)  
Options  
:
Large size adult arm cuff, adult arm cuff, small size adult arm cuff, cuff cloth  
of each size, air bag of each size, air tube (1.3 m), air tube (1 m), battery pack,  
stand exclusive for this unit, wall-hanging kit, pole-mounting kit  
* Please understand that speci[cations may be changed without prior notice.  
= Class II  
= Type B  
This blood pressure monitor ful[ls the requirements of the EC directive 93/42 EC directive 93/42/EEC (Medical Device  
Directive). It also conforms to the European standard EN 1060, Non-invasive Sphygmomanometers. Part 1: General  
Requirements an Part 3: Additional Requirements for Electromechanical Blood Pressure Measuring Systems.  
24  
 
Manufacturer  
OMRON HEALTHCARE Co., Ltd.  
53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN  
EU-representative OMRON HEALTHCARE EUROPE B.V.  
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS  
EC  
REP  
OMRON HEALTHCARE Co., Ltd.  
Production facility  
Subsidiary  
Mie, JAPAN  
OMRON HEALTHCARE UK LTD.  
Opal Drive  
Fox Milne, Milton Keynes, MK15 0DG, U.K.  
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH  
John-Deere-Str. 81a, 68163 Mannheim, GERMANY  
OMRON SANTÉ FRANCE SAS  
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE  
Made in Japan  
25  
 
English  
Important information regarding Electro Magnetic Compatibility (EMC)  
With the increased number of electronic devices such as PC.s and mobile  
(cellular) telephones, medical devices in use may be susceptible to  
electromagnetic interference from other devices. Electromagnetic interfer-  
ence may result in incorrect operation of the medical device and create a  
potentially unsafe situation.  
Medical devices should also not interfere with other devices.  
Important information for users  
In order to regulate the requirements for EMC (Electro Magnetic Compat-  
ibility) with the aim to prevent unsafe product situations, the EN60601-1-2  
standard has been implemented. This standard de[nes the levels of  
immunity to electromagnetic interferences as well as maximum levels of  
electromagnetic emissions for medical devices.  
This medical device manufactured by OMRON Healthcare conforms to this  
EN60601-1-2:2007 standard for both immunity and emissions.  
Nevertheless, special precautions need to be observed:  
Do not use mobile (cellular) telephones and other devices, which  
generate strong electrical or electromagnetic [elds, near the medical  
device. This may result in incorrect operation of the unit and create a  
potentially unsafe situation. Recommendation is to keep a minimum  
distance of 7 m. Verify correct operation of the device in case the  
distance is shorter.  
Further documentation in accordance with EN60601-1-2:2007 is available  
at OMRON Healthcare Europe at the address mentioned in this instruction  
manual.  
Correct Disposal of This Product  
(Waste Electrical & Electronic Equipment)  
This marking shown on the product or its literature, indicates  
that it should not be disposed with other household wastes  
at the end of its working life. To prevent possible harm to the  
environment or human health from uncontrolled waste disposal, please  
separate this from other types of wastes and recycle it responsibly to  
promote the sustainable reuse of material resources.  
Household users should contact either the retailer where they purchased  
this product, or their local government of[ce, for details of where and how  
they can take this item for environmentally safe recycling.  
Business users should contact their supplier and check the terms and  
conditions of the purchase contact. This product should not be mixed with  
other commercial wastes for disposal.  
This product does not contain any hazardous substances.  
Disposal of used batteries should be carried out in accordance with the  
national regulations for the disposal of batteries.  
EMC & WEEE  
26  
 

Milwaukee Drill 4004 20A User Manual
Musical Fidelity Power Supply V PSUII User Manual
NEC Network Router PathManager User Manual
Net Optics Switch N IDP POBPSX 001 N IDP POBPLX 002 User Manual
Nortel Networks Answering Machine P0911632 User Manual
Northern Industrial Tools Portable Generator 10000 BDG User Manual
NXP Semiconductors Switch PiP3209 R User Manual
Palsonic Flat Panel Television TFTV535WS User Manual
Panasonic Digital Camera AW PH350 User Manual
Panasonic Security Camera BLC230 User Manual